ATMOS - Pipette Accuracy Tester

Air-displacement pipettes need to be checked at regular intervals in order to comply with the ISO 8655-2 accuracy standard. The calibration verification requires very careful weighting with a laboratory scale. ATMOS revolutionizes the way you monitor pipette accuracy.

With ATMOS, you can everyday check in 30 seconds the accuracy of all your pipettes. This allows a continuous monitoring of performance between the usual calibration procedures.

ATMOS is measuring in dry conditions the exact volume of air displaced by the movement of the pipette's piston.

  • 3 Models:

    Atmos 1000: P20 to P1000 pipettes, min. calib. at 2µl

    Atmos Nano: P1 to P20 pipettes, min. calib. at 0.1µl

    Atmos Multi*: P20 to P1250, 8 channels pipettes
    (*available mid-2020)
  • Direct measurement, no need of liquid
  • Easy to use, compact (can be placed at the bench)
  • Pipette accuracy monitoring done in less than 30 seconds
  • ISO 8655-2 compliant pipette check protocol done in less than 3 minutes
  • Upload the calibration check report to a computer (USB connection)
  • Patent pending

Please contact us or our distributors for more information

Click below to access the Quick Guide and  a demo video :

ATMOS_1000_quick_guide1.03.pdf ATMOS_1000_quick_guide1.03.pdf
Size : 2319.436 Kb
Type : pdf
ATMOS_1000_manual_1_03.pdf ATMOS_1000_manual_1_03.pdf
Size : 3783.73 Kb
Type : pdf
video_3s1.mov video_3s1.mov
Size : 16721.352 Kb
Type : mov
clip_v2s.mov clip_v2s.mov
Size : 6021.354 Kb
Type : mov


FREQUENTLY ASKED QUESTIONS

1) Can I use ATMOS to calibrate my pipettes ?

The ISO 8655-2 norm for pipette calibration is based on gravimetry or dilution. ATMOS is an alternative method which is in principle as accurate (see note 3 below), but not yet part of the ISO 8655-2 standard. The best use of ATMOS for daily monitoring of the accuracy of your pipettes. It will detect any change of calibration that may, for example, be related to a shock (fall on the floor) or forcing the dial beyond the limit. This allows you to put damaged pipette on hold and not use them until your next campaign of calibration by gravimetry. It is of course also possible to use ATMOS to readjust a pipette in case of shift of accuracy (most of pipettes have a special tool to adjust the volume). This is much easier than gravimetry.

2) What is the leak measurement ?

ATMOS is measuring the air leak of the pipette or of the tip during its function. High leak can be related to a damaged piston in the pipette or to a defect in connection of the disposal tip. The leak affects the accuracy of the pipette, especially when using viscous liquid. Therefore, a pipette with a leak should be serviced (grease the piston and check tip cone connection). ATMOS will indicate wrong results when the pipette is leaky.

3) How ATMOS accuracy compares with gravimetry method in ISO 8655-2 standard ?

The random error of ATMOS readings are similar to those observed in gravimetry measurement with a laboratory scale.
The absolute accuracy of ATMOS stay within ½ of the ISO 8655-2 standard. ATMOS is measuring the true volumetric displacement of the piston in the pipette. When aspirating liquid in the pipette tip, the liquid height in the tip creates a negative pressure that is slightly lowering the effective volume of aspirated liquid. Therefore, ATMOS is using a compensation algorithm. The compensation algorithm is based on the use of standard disposable tips. When calibrating by gravimetry your pipettes with special tips, such as long tips, the ATMOS absolute value may differ.

4) Does the ATMOS need re-calibration service ?

ATMOS is measuring the volume dispensensed by the pipette by comparing with an internal fixed volume cavity. This comparison is done by measuring the pressure change when opening a valve towards a fixed reference volume. Since this reference volume has no moving parts and is made of stainless steel, it cannot change with time. Therefore, the ATMOS does not need re-calibration for it internal reference.
The pressure is measured by a precision pressure microsensor, which is factory calibrated. According to the long-term drift specifications of the sensor, we estimate that the ATMOS reading will stay within its specified accuracy during more than three years. Even after this time, we don’t anticipate substantial drift.
However, considering the stability concerns of our customers, we propose an instrument check service. If the ATMOS is sent to us, we will test it on our factory calibration bench and adjust the calibration value in the software when necessary. This service cost CHF 300.- (+ shipping and tax).

5) Is it possible to export data ?

ATMOS is designed to be compatible with automated platform and can be interfaced to any computer system (Window, Mac, Linux or other). Connection to the microUSB port allows a direct interface that can be used for pipette calibration. ATMOS microUSB is connected to the industry standard CP2102 from Silicon labs (refer to user's manual for more information).On request, we will send you an executable program ReadCalib (only compatible with Windows. When launching ReadCalib, it will be automatically recognized by your computer and you will be asked to enter a file path and name. All the ATMOS internal data are then stored at the chosen location in a Microsoft Excel format.



Distributors
Please contact UniPix if you are interested to distribute our products

Switzerland:

Life Systems Design - Starlab Switzerland
Pilatusstrasse 41, 5630 Muri, Switzerland
www.lifesystemsdesign.ch
Email

EUROPE:
UniPix direct sales

Israel:
Biological Industries Israel Beit Haemek Ltd.
Beit Haemek, Israel 25115
 
www.bioind.com
Email

China :
Leads Bio-Technology Limited
Hong Kong 999077, China
www.leads-tech.com
Email

Singapore:
Singlab Technologies Pte Ltd
www.singlab-biotech.com
Email

Korea:
to be announced